Technical Parameters

Model

CZ-UHA-
225G

CZ-UHA-
408G

CZ-UHA-
608G

CZ-UHA-
1000G

CZ-UHA-
1500G

CZ-UHA-
2000G

Internal Volume (L)

225

408

608

1000

1500

2000

Internal Size (W×H×D) (mm)

500×750×600

600×850×800

800×950×800

1000×1000×1000

1000×1200×1250

1200×1200×1400

External Size (W×H×D)(mm)

700×1785×1663

800×1975×1863

1000×2075×1863

1200×2125×2063

1200×2325×2313

1400×2325×2463

Approximately Weight (Kg)

520 

640 

710 

825 

1200 

1500 

Temperature Range

-70℃ ~ +150℃

Humidity Range

20%RH~98%RH(20~90℃, No load)

       
Temperature Deviation

±2℃ (No load)


Temperature Fluctuation

≤ ±0.3℃ (No load)

Temperature Uniformity

≤ 2℃ (No load)

Heating Rate

≥ 3℃/min (Average,No load)

Cooling Rate

≥ 1℃/min (Average,No load)


Humidity Deviation

+2%RH, -3%RH(Humidity > 75%RH), ±5%RH(Humidity ≤75%RH)


Humidity Fluctuations

±3%RH


Controller


7-inch 1024×600 dot matrix TFT color LCD display


Communication

   
Standard RS232 interface and RJ-45 network interface,with localand remote communication functions, can achieve WEB remote control


Auxiliary Functions


Fault alarm and cause,power-off protection function, calendar timing function(automatic start and stop operation),self diagnostic function

Function Configuration

Protective Device
● Non fuse switch
● Compressor overload protection switch
● High and low pressure protection switch for refrigerant
● Over humidity and over temperature protection switch
● Leakage protection,total power overload, short circuit protection
● Fault warning system
Accessories
● Silicone plug 1 piece
● Observation window
● 50mm test hole
● Lighting fixtures
● Video surveillance system(Purchase)
● APP mobile platform management(Purchase)

Meets the Test Standards

● GB/T 2423.1-2016 Low temperature test method Bb
● GB/T 2423.2-2008 High temperature test method Ab
● GB/T 2423.3-2016 Test Ca: Steady damp heat test method
● GB/T 2423.4-2008 Test Db: Alternating damp heat method
● GB/T 2424.5-2021 Environmental testing for electrical and electronic products Temperature test chamber performance confirmation
● GB/T 2424.6-2021 Environmental testing for electrical and electronic products Temperature and humidity test chamber performance confirmation
● GB/T 2424.7-2006 Environmental testing for electrical and electronic products TestA and B(with load)temperature test chamber measurement
● GJB 150.3A-2009  High temperature test
● GJB 150.4A-2009  Low temperature test
● GJB 150.9A-2009  Wet heat test

TAG: