Home/Applications/Heatsink contact pressure, read at the joint
Application 04 · Thermal interface

Heatsink contact pressure, read at the joint

A heatsink is rigid and a die is flat, so the rated mounting torque can still leave most of the interface with a gap. The pressure that decides thermal resistance stays invisible — unless a film goes into the joint and measures it.

0.3 mm piezoresistive film low-range array 20 Hz / 12 bit read during assembly
The measurement problem

Mounting torque is a force, not a field

An active device dissipates heat into its package, the package into a heatsink. For that path to work the two surfaces have to close across their whole area, at a pressure high enough to keep the thermal interface material at its design thickness. The difficulty is mechanical: a heatsink base is stiff, the mating surface is hard, and neither is perfectly flat. Clamp them together and contact usually starts at one or two points and leaves the rest of the nominal footprint with a gap.

A torque wrench controls the total clamping force. It says nothing about how that force divides across the base. Two heatsinks fastened to the same torque can differ by a factor of several in mean interface pressure, and the one with the higher peak can be the one with the worse thermal path — because the peak is carrying the load while the rest of the surface is doing nothing.

The clamp load is what you applied. The contact field is what the die actually sees.
01

Base flatness and warp

A base that is convex by a few tens of microns closes at the centre and leaves the perimeter open. Torque is on target; the perimeter is not conducting.

02

Fastening sequence

Four screws tightened in the wrong order tilt the base. The last screw to be tightened carries a disproportionate share, and the opposite corner loses contact.

03

Interface material thickness

A pad that is thicker in one region compresses the contact there and relieves it elsewhere. The map shows which region paid for it.

Method

A film thin enough to stay in the stack

The array is 0.3 mm thick with a 1.5 mm sealed border and a 300 mm lead. It is laid on the heatsink base or on the heat source, the assembly is closed at the torque and sequence used in production, and the pressure field is captured while the joint is closed. After the reading the film comes out and the thermal interface material goes in.

MEASURE

During assembly, not after

The map is read while the joint is closed, at production torque and production sequence. A slow, careful closure flattens exactly the gradient the check is looking for.

OPTIMISE

Turn a judgement call into a number

Screw order, washer size, base thickness, pad hardness — each change moves the field in a way the map records. The comparison is the deliverable, not the single picture.

SCREEN

Sample a batch

The same measurement on a sample from each batch separates units that are within the process window from units that are not, before they are built into a module.

TRACE

Keep the record

Every reading is stored with its seal, peak, mean and standard deviation, so a process change three months later can be judged against the numbers rather than against memory.

What the map showed

A 2,505 mm² footprint, half of it doing nothing

The trace in the gallery above is a real heatsink assembly, not an illustration. Across a 2,505 mm² footprint the mean was 163 kPa and the peak 1,000 kPa, with a standard deviation of 203 kPa — a spread larger than the mean itself. Total force was 409 N, which on the press dial looks entirely reasonable.

The field explains why: contact concentrated at both ends, where the fasteners sit, and dropped away through the middle. That is the signature of a base that is not flat under load, or of a fastening sequence that tilts it. Neither is visible from the torque value, and neither would be found by checking the joint temperature alone — the die would simply run hotter than the model predicted, and the difference would be attributed to the interface material.

ParameterMeasured by this setupWhat it means in this interface
Contact area read2,505 mm²The area that actually closes — not the area of the heatsink base.
Mean pressure163 kPaWhat the thermal interface material sees on average, which sets its compressed thickness.
Peak pressure1,000 kPaThe fastener-side ridge. The ratio of peak to mean is the measure of how uneven the joint is.
Standard deviation203 kPaHigher than the mean. In a well-designed joint this number is a fraction of the mean, not a multiple of it.
Total force409 NWhat the press or the torque calculation reports. It is correct, and it is not enough.

One production assembly, one torque setting. The numbers are reported as measured; they describe this joint, not a specification.

Measured

What the array saw

Traces from real assemblies, with the numbers read at the time. Nothing in this gallery is a rendered illustration.

Sensing array laid between a heatsink and its mounting plate, connected to a data collector and a laptop
Test setup: the array sits between the heatsink base and the heat source, one collector feeds the laptop.
Pressure mapping software showing a rectangular heatsink footprint with high pressure at both ends and a low-pressure band across the middle
A real trace from a 2,505 mm² footprint: mean 163 kPa, peak 1,000 kPa, standard deviation 203 kPa. High at both ends, starved across the middle.
Two-dimensional pressure map of a heatsink footprint
The same joint as a 2D field. The gradient is continuous, so the shape of the error can be read, not just its location.
Three-dimensional pressure surface of a heatsink footprint
Rendered as a surface: the two end ridges are the fastening points, the valley between them is where the interface never closed.
Pressure indicating film after clamping, showing a coloured footprint of the contact
The same interface on pressure-indicating film. One peak snapshot, no numbers, no second reading — fine for a go/no-go check, not for a tolerance.
Honest limits

Where this setup stops working

Every one of these has cost somebody a re-test. They are listed here rather than discovered on site.

This needs a low-pressure range

The standard PD-FS build covers 400–6000 kPa. Heatsink interfaces usually sit well below that, so the array for this work is a low-range build — the trace above was taken on a 0–400 kPa range. It has to be specified as its own item, with its own calibration, not assumed from the standard sheet.

The film occupies the interface it measures

At 0.3 mm it is the same order as the thermal interface material itself. Used as a check it is removed afterwards; it is not a layer you leave in a shipped assembly. Where the film is left in for a repeat reading, the compressed thickness of the interface changes and the thermal result changes with it.

Room temperature only

Readings are taken with the assembly at ambient. At operating temperature the heatsink, the die and the fastener stack expand differently, and the contact field moves — usually toward more contact at the centre and less at the edge. This method characterises the cold joint; the hot joint has to be inferred or measured another way.

20 Hz is for assembly, not vibration

Press closure, screw tightening and creep under load are all inside the sampling window. Shock, transport vibration and fan-induced excitation are not. Those need a different acquisition chain and a different sensor.

Surface condition reads as pressure error

A contaminated base, a thick conformal coating or a damaged pad surface changes the contact mechanically. The map will show it as low pressure, which is honest but easy to misread as a flatness problem.

It does not measure thermal resistance

The map shows the mechanical cause. It does not report junction temperature or thermal impedance. Correlating the two is a separate experiment, and a useful one — but it is a correlation, not a conversion.

FAQ

Frequently asked questions

No. The film is a compliant polymer layer with a sealed border; it is placed between the two surfaces and removed after the reading. What it does require is a clean surface on both sides — contamination is the most common reason a first reading looks wrong.

Not in the standard build. It is 0.3 mm of polymer in a thermal path, and it would change the interface thickness it is meant to be checking. It is a process-development and inspection tool: measure, remove, assemble with the intended interface material.

Usually far below the standard film range. The assembly in the trace above ran on a 0–400 kPa array. State the expected mean and peak at the enquiry stage — the low-range build is a separate item with its own calibration, and it is not something to extrapolate from a 400–6000 kPa sheet.

The standard sheet is 4096 points on 490 × 520 mm, which is far too coarse for a die footprint of a few square centimetres. Small-area work uses a custom array geometry at a much shorter pitch. Send the footprint dimensions and the smallest feature you need to resolve.

It answers a different question. Indicating film gives one irreversible peak snapshot, with no numbers and no repeat. The electronic array gives a calibrated field you can read again, compare between builds, and store. Film is cheaper and thinner; if a go/no-go check is all you need, film is the right tool.

For sampling, yes — the reading takes the same time as closing the assembly. For 100% inspection it is not the usual answer, because the film has to be handled, inserted and removed at line speed. Most users apply it to process qualification and batch sampling, not to every unit.

Send us the surface and the decision it has to support

Area, peak pressure, the feature you are looking for, and what happens if you miss it. That is enough for a first assessment.